Defibrillator Testers DEFTER 110

Equipped with a digital storage oscilloscope to display the waveform of biphasic shock energy discharge.

Analyzes key device parameters during operation, including the discharged energy for biphasic, monophasic, AED(manual mode), and multi-pulse shocks.

Can be connected to a computer to send and print test reports for the device under test.

Can be connected to probes for measuring current and voltage and sending data to the test device.

Load: 50 Ω ± 1% (25, 100 Ω Optional), with inductance < 70 μH

Resolution: 0.1 J  / Range: 0 J to 1000 J  / Pulse Width: 1ms to 50ms / Maximum Current: 110A Maximum Voltage: 5500V  / Minimum Voltage: 20V

Accuracy: 1000 J Range: ±2% of reading; 100 J to 1000 J: ±2 J; 100 J Range ±2% of Reading; ±0.1 J Waveform Storage: Discharge viewable

Voltage Accuracy: 1000 J Range: ±10V; 100 J Range: ±25 V / Current Accuracy: ±1

RS232 Output / Computer Control

 Computer control allows the user to operate the DEFTER 110 remotely via a serial RS232 interface. It requires an RS232 cable and a bidirectional D-9 connector

General Information

Display: 64*128 Pixel Graphic LCD Power: 12V Rechargable battery (Li-ion)

Dimensions: 32 cm (L) * 24 cm (W) * 10 cm (H)  / Weight: 3.8 Kg

 Safety Requirements : IEC/EN 60601-1

IEC 61000-4-2:2006 / IEC 61000-4-3:2012/ IEC 61000-4-4:2012 / IEC 61000-4-5:2014 / IEC 61000-4-6-2013  IEC 61000-4-8:2009 / IEC 61000-4-11:2004 /IEC 61000-3-2:2014 / IEC 61000-3-3:2013

EN 55022-2010 Class A /  EN 55022-2010 ClassB

Regular testing of the electroshock device at specific intervals ensures its proper functioning in emergency situations. The most reliable and comprehensive method for testing an electroshock device is to discharge it into a 50-ohm artificial load. For this purpose, Moj Parvaz Tasvir Company has designed and introduced DEFTER 110 in two models: 1. Fixed 50Ω load and 2. Variable load (25, 50, and 100 Ω) to test electroshock devices in a comprehensive manner.
ECG Waveforms : Rate: 60 BPM / Rate Accuracy: ± 1x of setting / Amplitude: Fixed at 1 mV lead II (RA-LL); fixed at 1.1 mV (Apex-Sternum)

Advantages of an electroshock tester

Easy to use

portable

synchronize test

 

Ability to measure the output energy of all monophasic, biphasic, and AED(manual mode) shocks from minimum to maximum joules

 

Ability to connect to a computer and print output tables for use in hospital PM program

10-year after-sales service and a 1-year warranty

 

Parameters provided by the device

The device displays the output waveform in 20ms and 10ms intervals

Ability to measure the full range of output energy, from minimum to maximum joules

Ability to measure peak voltage in all shocks, displaying both positive and negative voltage values in biphasic shocks

Ability to measure peak current in monophasic shocks

-Ability to measure T+ , T

Ability to measure the R-wave duration and sync test