Equipped with a digital storage oscilloscope to display the waveform of biphasic shock energy discharge.
Analyzes key device parameters during operation, including the discharged energy for biphasic, monophasic, AED(manual mode), and multi-pulse shocks.
Can be connected to a computer to send and print test reports for the device under test.
Can be connected to probes for measuring current and voltage and sending data to the test device.
Load: 50 Ω ± 1% (25, 100 Ω Optional), with inductance < 70 μH
Resolution: 0.1 J / Range: 0 J to 1000 J / Pulse Width: 1ms to 50ms / Maximum Current: 110A Maximum Voltage: 5500V / Minimum Voltage: 20V
Accuracy: 1000 J Range: ±2% of reading; 100 J to 1000 J: ±2 J; 100 J Range ±2% of Reading; ±0.1 J Waveform Storage: Discharge viewable
Voltage Accuracy: 1000 J Range: ±10V; 100 J Range: ±25 V / Current Accuracy: ±1
RS232 Output / Computer Control
Computer control allows the user to operate the DEFTER 110 remotely via a serial RS232 interface. It requires an RS232 cable and a bidirectional D-9 connector
General Information
Display: 64*128 Pixel Graphic LCD Power: 12V Rechargable battery (Li-ion)
Dimensions: 32 cm (L) * 24 cm (W) * 10 cm (H) / Weight: 3.8 Kg
Safety Requirements : IEC/EN 60601-1
IEC 61000-4-2:2006 / IEC 61000-4-3:2012/ IEC 61000-4-4:2012 / IEC 61000-4-5:2014 / IEC 61000-4-6-2013 IEC 61000-4-8:2009 / IEC 61000-4-11:2004 /IEC 61000-3-2:2014 / IEC 61000-3-3:2013
EN 55022-2010 Class A / EN 55022-2010 ClassB